Order # 29A02, Electric Sterilizer,  $ 770.00



New, moved or repaired automatic sterilizers are not considered ready to be put into service until three (3) consecutive "negative" spore test are received.
3 Free Pre-Paid spore tests are include with these units.
After the initial testing check with your HD for
their regulations regarding periodic testing. Order more spore tests.

The "Electric Processor," will require a "negative" spore test for every load before the items in that load are put into service.

All autoclaves that are not fully automatic
MUST spore test every load.
Items processed with a new unit cannot be used
until 3 "negative growth" spore tests are returned.
3 Free Pre-Paid spore tests are include with this unit.
Do not put this unit into service without running 3 consecutive loads with spore tests,
returning the spore tests for processing
and waiting for the results before putting the items into service.
After the initial testing
spore test every load.
Order more spore tests.

The Electric Sterilizer with built-in heater and temperature setting dial.  No Timer.
It should also be noted that metal-to-metal seal (no gaskets) the tops must be uniformly tightened working around and around little by little to prevent warping.

Large Electric Autoclave Steam Sterilizer # 29A-02

Description
Overall height 16 3/4" (25 Qt. 23.66 Liters)
Weight 24 Lbs
Maximum Pressure 20 psi
Maximum temperature 259 Degrees F Calibrated Gauge
Self-contained electric heat source
Steam relesae switch.
Voltage 110V
Chamber Size: 11" Diameter x 10" Deep
Cylce time from cold 45 minutes.
Time to monitor is manual, no automatic shut off, manual type only.